Vue de la page 35
32
- Within the Rate menu
Per: Set the period between pulses.
- Within the Gate/Trig menu
EXTin: Select Triggered mode.
Level: Set the trigger threshold level.
Edge: Select which edge, rising or falling, to trigger on.
Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will
generate a set of pulses for every external trigger received. The unit is reset at
the end of a burst and will generate another set of pulses upon receiving a new
trigger. Triggers that occur in the middle of a burst are ignored. Pressing the
RUN/STOP key will disarm the unit.
To Start a Continuous Stream of Pulses Using the External Trigger
The external trigger may be used to start the unit generating pulses:
- Within the Mode menu
Mode: Select the Continuous mode.
- Within the Rate menu
Per: Set the period between pulses.
- Within the Gate/Trig menu
EXTin: Select the Trigger mode.
Level: Set the trigger threshold level.
Edge: Select which edge, rising or falling, to trigger on.
Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will
begin generating pulses after an external trigger is received. Triggers that occur
after the pulses start are ignored. Pressing the RUN/STOP key a second time
will disarm the unit.
To use the External Gate to Control the System
The external trigger may be used to control the output of the unit. To gate the
system timer:
- Within the Mode menu
Mode: Select the desired mode.
- Within the Rate menu
Per: Set the period between pulses.