10-39
System Commands
Keyword
Command
Sub
-
Command
Sub
-
Command
Parameter
Range
Notes
:SYSTem
:STATe ?
Query only. Returns ”1” if the system
is armed and/or generating pulses or
“0” if the system has been disarmed.
:BEEPer Controls the audible beeper.
:STATe
0/1 or
OFF/ON
Enables/disables the beeper.
:VOLume 0-100
Sets the volume of the beeper. 0 is
off and100 is maximum volume
:COMMunicate
Controls the system communication
parameters
:SERial
Controls the serial parameters.
:BAUD
4800 / 9600 /
19200 /
38400 /
57600 /
115200
Sets the baud rate for both receiving
and transmitting using the RS232
port.
:USB
4800 / 9600 /
19200 /
38400 /
57600 /
115200
Sets the baud rate for
communication when using virtual
com ports for USB.
:ECHo
0/1 or
OFF/ON
Enables/Disables transmission of
characters received on the RS232
port.
:INTLCK ?
Query only: Returns the current state
of the Arm Switch/Interlock circuit.
:KLOCk
0/1 or
OFF/ON
Locks the keypad.
:VERSion ?
Query only. Returns SCPI version
number in the form: YYYY.V ex.
1999.0
:BVERsion ?
Query only. Returns the current
version of the bootloader.
:INFOrmation ?
Query Only. Returns model and
version information for the
instrument.
:NSID ?
Query Only. Returns the instrument
ID for use when upgrading firmware.