9530/8530 Manual Version 4.8 Page 17
Setting System Mode of Operation
By pressing the Menu button the system mode menu can be accessed. The system Mode menu
sets the T
0
system timer mode. The mode may be changed by setting the cursor on the mode row
and pressing the up and down key. By pressing the next button while in this menu will allow the
user to change between setting the mode or the parameter. The menu will show the extra set of
parameters (Brst, On & Off) only when they are applicable.
T
0
Mode:
Parameters:
Continuous
N/A
Single Shot
N/A
Burst
Brst:
‘N’ number of pulses to generated for each T
0
pulse
Duty Cycle
On/Off:
‘N’ number of pulses to generated for each T
0
pulse / ‘M’ number of pulses to inhibit for each
T
0
pulse
*NOTE: Any mode may be started by either the RUN/STOP key in the internal trigger mode or
armed by the RUN/STOP key and started by an external trigger in the external trigger
mode. In the single shot and burst modes, (internally triggered) the unit disarms itself
at the end of the pulse train. Pressing the RUN/STOP key after the unit has been
disarmed will generate a new pulse train.
Clock/Rate Menu
T
0
Rate:
Source: Ext Osc
Rate: 100 MHz
Level: 0.50 Volts
Ref Out: T
0
Pulse
Source: Int Source
Press the Page Button
Change Clock Sources
Ref Out: T
0
Pulse
Level: 0.50 Volts
Source: Ext w/ PLL
Rate: 100 MHz
Level: 0.50 Volts
Ref Out: T
0
Pulse
Ext Opt: Force Off
Level: 0.50 Volts
T
0
Rate: T
0
Rate: