46
Keyword
Lower-Level
Keyword
Parameter
Std /
New
Comments
:DISPlay
Std
Subsystem. Contains commands to control the display.
:MODe 0 / 1 or OFF / ON New
Enables / Disables automatic display update. When
true, front panel display is updated with serial command
parameter changes. Setting to false (0) decreases
response time.
:UPDate? New
Query only. Forces update of display. Use when mode
is false.
:BRIGhtness 0 - 4 New
Controls intensity of display. Range is 0 to 4, where 0 is
off and 4 is full intensity.
NOTE: DISPlay commands are considered irrelevant for the 8510 series.
IEEE 488.2 Common Commands
Mnemonic Command Name Parameter Comments
*IDN? Identification Query
Queries the pulse generator identification. The ID will be in
the following format: model# - #channels - option# -
version#
*RCL Recall Command 0 - 12
Restores the state of the pulse generator from a copy stored
in local nonvolatile memory (0 through 12 are valid
memory blocks).
*RST Reset Command Resets the pulse generator to the default state.
*SAV Save Command 1 - 12
Stores the current state of the pulse generator in local
nonvolatile memory (1 through 12 are valid memory
blocks).
*TRG Trigger
Generates a software trigger pulse. Operation is the same as
receiving an external trigger pulse.
*LBL Setup Label ? or "string value"
The <?> is used to query the label of the last saved or
recalled configuration.
The string must be in double quotes and no longer than 14
characters. Command must be followed by a *sav [1 / 2 / n]
command to take effect. Note: to see the label on the screen
a display update or reboot must take place